The interpretation of the calibration certificate is essential for proper decision making
The correct interpretation of the calibration certificate is necessary for making decisions regarding the validity of the equipment or standard and regarding the adjustments that are subsequently made.
In this entry, we will explain what a calibration procedure entails, as well as the difference between this procedure and the zeroing procedure for the equipment. We will show an example of calibration of a spectrophotometer for color measurement. We will also discuss metrological traceability and the verification of measuring equipment. Finally, we will focus on the interpretation of calibration certificates and the calibration plan.
Calibration of the measuring equipment or standard
It is recommended to have the calibration of the instrumental equipment performed by a laboratory accredited by ENAC. Said laboratory will carry out the calibration on standards of certified value and will issue a report that will collect the deviations of the equipment in taking measurements.

A calibration is a set of operations that determine the relationship between the values given by a piece of equipment and the nominal value of a standard for a given magnitude.
Thus, the main task of calibration is to guarantee the veracity of the results obtained, detecting any systematic error in the measurement and the precision to control the variability of the results due to problems in the repeatability of the measurement.
Traceability
When calibrating a piece of equipment, it is important to bear in mind that it must guarantee the traceability of the measurements to be carried out. In fact, the norm ISO 9001 obliges companies to ensure traceability, and defines it as follows:
“Traceability is the property of the result of a measurement or the value of a standard where it can be related to specified references, usually national or international standards, through a continuous chain of comparisons all with specified uncertainties.” (ISO 9001. International Vocabulary of Basic and General Terms in Metrology).
However, internal or external methods can be defined to fulfill this requirement. In the case of performing it externally, it is advisable to work with an accredited laboratory.
For its part, ENAC at their metrological traceability policy (NT-74), specifies that the conformity assessment bodies (OEC) must guarantee that they maintain the metrological traceability of the measurements, for which they must ensure that said measurements are part of an uninterrupted chain of comparisons, originating from adequate references.
Traceable calibration and accredited calibration
A traceable calibration can be carried out by a laboratory through a standard or reference equipment that must guarantee traceability. It is recommended in traceable calibrations that the laboratory provide the traceability of the standards or types of reference used (calibration certificate).
For its part, a calibration accredited by ENAC This type of calibration is performed by an accredited laboratory. The procedure involves determining the relationship between the values provided by the measuring equipment and the normal value relative to a standard. Traceability in measurement is implicit in this type of calibration. Thus, after performing the calibration, the laboratory will issue a certificate bearing the ENAC mark, guaranteeing traceability.
Calibration Example
The calibration of colorimeters and spectrophotometers determines, under specified conditions, the relationship of the values of the measured chromatic coordinates and the corresponding values of those coordinates in comparison with national and international metrological traceability standards.
C= (Vmeasured - Vcertificates) ± U
Where U = expanded uncertainty
Below we show an example of calibrating a spectrophotometer for color measurement.

The conditions of habitual use of the equipment must be taken into account: geometry, illuminant and standard observer.
Other concepts
Zero adjustment / fine tunning
Before taking measurements with the equipment, it must be adjusted to zero, which is carried out on its own adjustment piece supplied by the manufacturer, following its instructions.

It is important that this adjustment piece is kept in perfect condition, without breakage, blows or dirt that may interfere with the results. In addition, it is necessary to know that these pieces are not interchangeable, that is, each measurement equipment is supplied with its own piece, and this must be sent to the laboratory together with the equipment and the connection wiring in order to carry out the calibration.
The zero adjustment allows to ensure the operation of the equipment according to the manufacturer's specifications.
Verification
The verification of a measurement equipment consists of taking the results of the calibration and uncertainties and ensuring that the equipment complies with the specifications established by the manufacturer or the company. In other words, it is a theoretical exercise designed to be sure that no requirement is unfulfilled in the design.
Therefore, verification is not the same as a calibration, traceable or accredited.
Equipment adjustment and recalibration
After calibrating a piece of equipment, a systematic error may be detected. In this case, it is possible to adjust the measuring equipment, making the indication of the calibrated instrument as close as possible to that indicated by the standard.
Once the adjustment has been made, a recalibration of the equipment must be carried out.
However, not all measurement instruments can be adjusted, nor can all systematic errors be corrected or systematized.
Interpretation of the certificate
A calibration aims check whether a product or piece of equipment meets a given specification, that is, if the values that have been obtained for one or several characteristics of said product or equipment are within the limits that have been established as tolerated.
As we have said previously, the interpretation of calibration certificates must be carried out properly. It is essential to understand the content of the report issued by the accredited laboratory and know how to interpret the data to make the right decisions about the equipment and its use.

To carry out the calibration, the laboratory applies a rigorous procedure for taking measurements and calculating the correction values and uncertainty, concepts that are important to understand and that we will explain below.
What should a calibration certificate include?
A calibration certificate carried out by a laboratory must include the results obtained in an exact, clear, objective and unambiguous manner.
Specifically, the certificate must include:
- Title ("Calibration certificate").
- Name and address of the laboratory.
- Place where the activities are carried out.
- Unique identification of all components of the certificate (for example, paging).
- Customer name and contact information.
- Identification of the method used. When the method is standardized or reference is made to a standardized method, the revision status or year of approval of the document that identifies said standardized method must be included.
- Dates of execution of the laboratory activity.
- Date of issue of the certificate.
- Statement that the results relate only to the items undergoing calibration.
- Results with units of measurement.
- Method additions, deviations, or exclusions.
- Identification of the persons authorizing the certificate.
- Clear identification when the results come from external providers.
- Appropriate use of the ENAC mark on the certificate.
- Measurement uncertainty of the result.
- Conditions in which the calibrations have been made, for example, environmental conditions.
- Statement identifying how the measurements are metrologically traceable.
- Results before and after adjustment or repair if available.
- May include calibration interval recommendations, if indicated by the customer.
Correction
In order to analyze the behavior of the equipment, we need to know what the correction is for each parameter or measurement quantity. The correction is the difference between the value measured by the equipment to be calibrated on each stndard and its certified value.
Thus, the information related to the correction that appears in the certificate is useful to correct the reading offered by a device during its daily use.
Following the spectrophotometer calibration example, the corrections on each coordinate indicated in the report will be those applied to future measurements, in addition or subtraction, depending on the correction, to the measurement value. The technician in charge of measuring with the equipment in the company must apply the corrections provided in the report for each coordinate.
Uncertainty
In every measurement process there is an uncertainty that must be taken into account, which depends on various factors such as, for example, the methods of the measurement process, the equipment used for it, the facilities themselves and environmental conditions, the manipulation of test items, human factors, etc.
Thus, the uncertainty includes all those sources or contributions of error that affect the calibration and its value is obtained from calculation formulas.
Generally, the certificates indicate the expanded uncertainty that has been obtained by multiplying the typical measurement uncertainty by the coverage factor k=2 which, for a normal distribution, corresponds to a coverage probability of approximately 95%.
Calibration plan
The industry uses a large amount of measuring equipment to ensure the quality of the products it produces. Therefore, to guarantee that these measurements are reliable, that is, that they are traceable and, therefore, can be related to national or international standards, it is convenient that the equipment be calibrated periodically. For this, companies should have implemented a calibration plan as a fundamental part of the quality plan.
Thus, the activities of the calibration plan will have the objective of detecting possible errors or malfunctions of the equipment.

The calibration plan has to define the calibrations that are necessary and detail if they are internal or external.
If you choose to carry out calibrations internally, a series of procedures to be applied must be established. In addition, you have to make sure that the technical employees who perform the calibrations have the appropriate training. It is recommended that these points be included in the calibration plan, as well as if the company considers the acquisition of new instruments or standards necessary.
The calibration plan contains the identification of all equipment, standards or measuring instruments and accessory elements for which suitable calibration intervals are methodically established. The purpose of this is to ensure that the measurements carried out with these devices follow the traceability chain.
In addition, whether the company chooses to perform the calibrations internally or externally or both, it is always recommended that they carry out an internal control between calibrations.
On the other hand, if the calibrations of the equipment are carried out externally, it will be an accredited laboratory that, with a periodicity stipulated by the company, will calibrate the equipment and issue a certificate with the results of the calibration. It is recommended that the company detail in the calibration plan the selection of the external laboratories that best meet the criteria it deems.
Laboratorios Eyco
Laboratorios Eyco has the largest scope of ENAC accreditation in Spain in the optical area. In addition, Eyco also offers advice for the preparation of the calibration plan for optical equipment and standards.
There are many companies, laboratories and entities that trust in our experience and competence. As we have previously indicated, the ENAC accreditation guarantees the traceability of the results. The reports or certificates issued under ENAC accreditation are recognized by the signatories of the EA, ILAC and IAF agreements, thanks to the international agreements of the accreditation bodies. For this reason, companies and laboratories around the world already trust Laboratorios Eyco.
FAQ
What is measurement uncertainty in a calibration certificate?
La measurement uncertainty It is the parameter that quantifies the dispersion of the values that could reasonably be attributed to the measured quantity. In a calibration certificate, it represents the technical margin of doubt of the process and is a fundamental piece of information for evaluating whether the equipment is suitable for its intended use within the company's tolerances.
What is the difference between a systematic error and its correction?
El systematic error It is the constant deviation that the equipment exhibits with respect to the reference or standard value. correction It is the numerical value that must be algebraically added to the uncorrected result to compensate for this instrumental deviation, thus guaranteeing technical rigor and accuracy in industrial measurement processes.
Why doesn't a zero adjustment replace a calibration?
El zero adjust It is a maintenance or optimization operation of the instrument so that it begins reading at a specific baseline point, but it does not provide information about the equipment's behavior across its entire measurement range. calibrationInstead, it formally compares the equipment readings with reference patterns to determine errors and uncertainty at different critical points.
What are the advantages of having a calibration certificate accredited by ENAC?
A certificate issued under the accreditation of ENAC (or international equivalents) It guarantees direct metrological traceability to national or international standards. It offers maximum legal and technical security in the event of quality management system audits, such as those under ISO 9001, formally demonstrating the laboratory's competence.
How often should a calibration plan be implemented in the company?
There is no single mandatory frequency; the periodicity of a periodic calibration plan The interval is established based on the criticality of the process, the frequency of instrument use, the manufacturer's recommendations, and the history of detected deviations. Defining optimal intervals allows for timely correction of deviations and continuous optimization of technical decision-making.