UV-Visible Spectrophotometer Calibrations
On-site or laboratory calibration, with a personalized and flexible service. Delivery time: 5 days.
What do we calibrate?
Calibration of UV-Visible spectrophotometers.
We offer 3 types of calibration:
STANDARD CALIBRATION
Recommended calibration for companies and laboratories that use the spectrophotometer on a daily basis in the UV-Visible range of 200-800 nm.
- Determining the accuracy of the wavelength at 5 points (if the equipment allows it): 279 nm, 361 nm, 453 nm, 536 nm and 637 nm.
- Determining the accuracy of the photometric scale:
UV range: 235nm, 257nm, 313nm and 350nm.
Visible range: 440nm, 465nm, 546.1nm, 590nm and 635nm.
- Determination of diffuse light at 340 nm and 370 nm.
- Determination of repeatability from the absorbance readings
PREMIUM CALIBRATION
Recommended for companies and laboratories that require a higher density of dots or specific dots within the UV-Visible range of 200–800 nm.
- Determining the accuracy of the wavelength at 5 points (if the equipment allows it): 279 nm, 361 nm, 453 nm, 536 nm and 637 nm.
- Determination of the accuracy of the photometric scale (absorbance/transmittance) in 12 points:
UV range: 235nm, 257nm, 313nm and 350nm.
Visible range8 points to be chosen between 420 nm and 900 nm. (If calibration points are not specified, the laboratory will calibrate at: 450, 500, 550, 600, 650, 700, 750 and 800 nm)
- Determination of diffuse light at 340 nm and 370 nm.
- Determination of repeatability from the absorbance readings.
CUSTOM TECHNICAL CALIBRATION
Recommended calibration for laboratories with their own protocols or equipment with special requirements.
- Determining the accuracy of the wavelength, from 250 nm to 2000 nm (if the equipment allows it).
- Determination of the accuracy of the photometric scale (absorbance/transmittance) at all points required by the client in the 200–2500 nm range. (If the points are not specified, the laboratory will calibrate the entire range of the device every 50 nm)
- Determination of diffuse light at 340 nm and 370 nm.
- Determination of repeatability from the wavelength and absorbance readings.
UV-Visible Spectrophotometers
UV-visible spectrophotometers allow the evaluation of the light that is absorbed, transmitted or reflected by a material – whether in a solid or liquid state – for each wavelength.
Transmittance is the fraction or percentage of incident light that passes through a sample at given wavelengths. Thus, by measuring transmittance, it is possible to determine whether a material transmits UV light, its transparency in the visible spectrum, its color (in translucent materials), and so on.
On the other hand, absorbance—also called optical density or optical transmittance (OD)—is the amount of light a material absorbs. This quantity is logarithmically related to transmittance. Absorbance is commonly used to establish a relationship, through the Lambert-Beer Law, with the concentration of a particular analyte being studied.
The measurements of these magnitudes are made through spectrophotometers. These equipments have mainly 3 components: the illumination system, the monochromator and the detector.

Main components of UV-visible spectrophotometers
Lighting system
The system may consist of different light sources. Spectrophotometers often have a hydrogen or deuterium light source for the ultraviolet spectrum and a tungsten light source for the visible and near-infrared (NIR) spectrum. Specifically, the light sources, which define the instrument's spectral range, can range from 200 nm to 2500 nm. UV-Visible spectrophotometers typically cover a wavelength range of 200 nm to 900 nm, encompassing the ultraviolet, visible, and near-infrared regions.
Monochrome
Furthermore, this component allows the light spectrum to be separated, thus providing an energy beam at a specific wavelength. This makes it possible to perform tasks requiring precise separation of different wavelengths. For example, if working with various chemical compounds that exhibit a characteristic absorbance at 340 nm, it is important that the monochromator precisely separates this wavelength from the others; therefore, it is essential to ensure that the equipment operates at 340 nm and not 342 nm.
Specifically, wavelength calibration is performed on this part of the equipment, especially when wavelength is critical or affects the measurements being taken. This type of calibration checks the difference between the wavelength measured by the equipment and the wavelength of the reference standard used. Specifically, it is possible to calibrate different wavelength points across the entire range of the equipment, from 200 nm to 2500 nm.
Detector and logger
Finally, the detector is responsible for transforming the signal it receives (photons) and converting it into an electrical signal, so that, depending on its intensity, it can be recorded by a system that provides the numerical values of transmittance and absorbance or DO.
Knowing the exact values of absorbance and transmittance at each wavelength is also a critical aspect, since it allows comparing spectra of different materials or solutions with different concentrations.
Thus, detector calibration, or the absorbance and transmittance scale, allows us to know the instrument's deviation in these values. Specifically, this is important because, for example, when measuring the absorbance of a material at 50 nm, an instrument might indicate that the material has an absorbance of 2 DO at this wavelength, when the actual value could be 2.3 DO. Furthermore, knowing the instrument's absorbance correction is necessary for performing accurate measurements.
Absorbance calibration is performed at all necessary wavelengths. Thus, measurements are taken for different absorbance values at each wavelength. The difference between the value measured by the instrument and the certified value of the standards used is then evaluated.
Other evaluable parameters: diffuse light or Straylight and the linearity of the photometric scale
On the other hand, stray light is radiation that can reach the detector at an unwanted wavelength. Specifically, this type of radiation can cause significant errors in absorbance readings and, consequently, deviations from the Lambert-Beer Law.
Finally, the linearity of the photometric scale involves determining the instrument's ability to measure absorbance as a function of the concentration of different solutions. Specifically, linearity can be analyzed using correlation coefficients of the absorbance calibration curve as a function of concentration.

Technical scope for the calibration of UV-visible spectrophotometers
Eyco is accredited by ENAC For the calibration of UV-Visible spectrophotometers. Specifically, on the wavelength scale and in absorbance or transmittance. Our laboratory is equipped to calibrate both portable and benchtop equipment.
Furthermore, Eyco is also accredited to calibrate the NIR region, from 780nm to 2500nm. Specifically, our laboratory is the only one in Spain accredited to calibrate the NIR region from 900nm to 2500nm.
| Measurement range Range | Uncertainty (*) -- | Standard/Procedure -- | Instruments to be calibrated Instruments | Dress Code |
|---|---|---|---|---|
| TRANSMITTANCE TRANSMITTANCE | ||||
| (213,235,257,261,313,350) nm 1 ≥ T > 0,3 0,3 ≥ T > 0,1 0,1 ≥ T > 0,01 0,01 ≥ T > 0,001 | 0,020 0,050 0,20 0,25 | PT39 Internal procedure based on: Royal Pharmacopoeia Spanish IUPAC Recommended Reference Materials for realization of Physicochemical properties | UV-VIS Spectrophotometers | A,I |
| 200 nm ≤ λ ≤ 420 nm 1 ≥ T > 0,3 0,3 ≥ T > 0,1 0,1 ≥ T > 0,01 0,01 ≥ T > 0,001 | 0,0090 0,015 0,040 0,15 | PT50 Internal procedure based on: Royal Pharmacopoeia Spanish IUPAC Recommended Reference Materials for realization of Physicochemical properties | UV-VIS Spectrophotometers | A,I |
| (420-900)nm 0.00010≤T<0.0010 0.0010≤T<0.10 0.10≤T±1 | 0.30 0.080 0.0069 | PT39 Internal procedure based on: Spanish Royal Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | UV-VIS Spectrophotometers | A,I |
| TRANSMITTING OPTICAL DENSITY (ABSORBANCE) OPTICAL TRANSMITTANCE DENSITY (ABSORBANCE) | ||||
| (213, 235, 257, 261, 313, 350) nm 0 ≤ A < 0,5 0,5 ≤ A < 1,0 1,0 ≤ A < 2,0 2,0 ≤ A ≤ 2,5 | 0,0070 0,0080 0,030 0,070 | PT39 Internal procedure based on: Spanish Royal Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | UV-VIS Spectrophotometers | AI |
| 200 nm ≤ λ ≤ 420 nm 0 ≤ A < 0,5 0,5 ≤ A < 1,0 1,0 ≤ A < 2,0 2,0 ≤ A ≤ 2,5 | 0,0040 0,0060 0,020 0,060 | PT50 Internal procedure based on: Royal Pharmacopoeia Spanish IUPAC Recommended Reference Materials for realization of Physicochemical properties | UV-VIS Spectrophotometers | AI |
| (420-900)nm 0≤A≤1.50 1.50≤A≤3 3≤A≤4 | 0.0040 0.050 0.20 | PT39 Internal procedure based on: Spanish Royal Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | UV-VIS Spectrophotometers | AI |
| (900-2500)nm 0≤A≤0.8 0.8≤A≤2 | 0.0040 0.0090 | PT39 Internal procedure based on: Spanish Royal Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | Spectrophotometers | AI |
| WAVELENGTH Waveength | ||||
| 200nm≤λ≤850nm 850nm≤λ≤2000nm | 0.30 nm 0.50 nm | PT39 Internal procedure based on: Spanish Royal Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | UV-VIS Spectrophotometers | AI |
The result
Once your equipment has been calibrated by Eyco's expert staff, you will receive a Calibration Certificate backed by accreditation. ENAC. Said Certificate includes the complete identification of your equipment, the procedure that the laboratory has followed to carry out the calibration and the result of the same.
The result of UV-Visible spectrophotometer calibration is a correction along with its uncertainty. This correction is the difference between the reference standard value and the value measured by your instrument. The Calibration Certificate issued by Eyco will include the wavelength and absorbance corrections. Upon request, stray light and linearity values will also be provided.
From the correction values and their uncertainty, you can establish the acceptance and rejection values of the equipment.
In addition, together with your equipment, you will receive a calibration label with the certificate number, the identification of the equipment and the date of calibration.