UV-Visible Spectrophotometers: Interpretation of Calibration Certificates

Absorbance/Transmittance

Understanding calibration certificates is key to making decisions regarding equipment

Once the calibration has been performed in UV-Vis spectrophotometers (ultraviolet-visible), accredited laboratories issue a certificate to the client. With the data provided by this report, the company can make decisions about the equipment and, where appropriate, about the adjustments that can be made to it and the measurements that the equipment is obtaining.

To do this, it is important to know how to properly interpret the data that appears in said calibration report. That is why in this entry, we will talk about key issues for the interpretation of a calibration certificate for this type of equipment, such as correction and uncertainty.

We will also refer to more general topics specific to the calibration process, and to concepts such as traceability, as well as the difference between traceable calibration and accredited calibration. We will also address issues such as zero adjustment, check, or the adjustment and recalibration of the equipment.

In addition, we will show examples of data obtained in the calibration of a UV-Visible spectrophotometer for different magnitudes, and we will explain what a calibration plan consists of.

UV-Vis spectrophotometers. Interpretation of the certificate.

The UV-Visible spectrophotometers They are fundamental tools in many laboratories in various sectors. These are devices that allow the measurement of the absorbance of substances depending on the wavelength of the incident light. Now, to ensure the precision and reliability of these measurements, it is essential to understand how to interpret the calibration certificates that accompany these instruments.

If the calibration has been contracted in wavelength (λ) and absorbance, the calibration certificate will show detailed results obtained for both wavelength and absorbance with all the wavelengths that have been contracted.

Spectrophotometer wavelength calibration provides data to ensure that the instrument is capable of emitting light at the required wavelengths accurately and that the absorbance or transmittance values ​​the equipment measures correspond to the correct wavelength.

Regarding absorbance calibration, it is crucial to validate that the spectrophotometer correctly measures the amount of light absorbed by a sample at different wavelengths. This calibration verifies that the absorbance readings are accurate and consistent with established standards.

On the other hand, diffuse light is radiation that can reach the detector with an unwanted wavelength. This type of radiation can cause significant errors in absorbance readings and, therefore, deviations from the Lambert-Beer law. If diffuse light is also measured during calibration, it is determining whether or not the equipment is receiving unwanted light.

In addition, the certificate may also show the values ​​relating to the photometric linearity of the equipment. That is, through a linear adjustment of the results, it is established how linear the team's behavior is. If it is linear, the equipment is behaving similarly throughout the analyzed wavelength range.

Calibration certificate
Image 1. Accredited calibration certificate

What should a calibration certificate include?

Calibration is carried out for the purpose of verifying the conformity of a product or equipment with certain predefined specifications. This involves ensuring that the values ​​obtained from the characteristics of the product or equipment are within the acceptable limits established according to its use.

UV-VIS Spectrophotometers
Image 2. UV-Vis spectrophotometer

The calibration certificate issued by a laboratory must present the results in a precise, clear, objective and unambiguous manner. These documents must include:

Correction/Error/Accuracy

To understand the performance of the equipment, it is essential to know the corrections associated with each parameter or magnitude of measurement. The correction represents the discrepancy between the value measured by the equipment to be calibrated and the certified value of the reference standards used.

The correction information provided in the certificate is essential for adjusting readings obtained during regular use of the equipment. For example, when calibrating a UV-Visible spectrophotometere, the corrections that appear in the report indicate the difference between the values ​​measured by the equipment and the certified values.

Uncertainty

La uncertainty in a measurement process refers to the lack of absolute accuracy or precision due to a variety of factors. These factors may include the methods used in the measurement, the equipment used, environmental conditions, handling of test objects and human factors, among others.

Uncertainty encompasses all possible sources of error that influence the calibration, and their value is determined by specific calculation formulas. Calibration certificates generally indicate the expanded uncertainty, which is obtained by multiplying the standard uncertainty of the measurement by a coverage factor, denoted as “k=2”. This factor, applied in a normal distribution, corresponds to a coverage probability of approximately 95%.

Wavelength

The UV-Vis spectrophotometers They have the ability to emit light at different wavelengths, and they use this property to identify and quantify different substances, since each compound absorbs light at specific wavelengths, creating a characteristic spectrum. By measuring the absorbance of light at different wavelengths, the spectrophotometer can determine the concentration of compounds present in a sample. This technique is widely used in chemistry, biology and materials science to analyze the composition and purity of various substances.

On a calibration certificate, wavelength data shows information about the correctness of the equipment. The correction is obtained from the difference between the certified value and the measured value, and has an associated uncertainty:

Thus, more specifically, for the calibration of the wavelength, standards that have absorbance peaks at specific wavelengths whose values ​​are certified. When measuring these standards with the equipment to be calibrated, the discrepancy between the reference values ​​and the values ​​of the absorbance peaks obtained with the spectrophotometerThus, in the correctionIt can be observed whether the equipment is capable of emitting light at the required wavelengths with the necessary precision for each case. The correction is always presented along with its associated expanded uncertainty.

Below is an example of how wavelength data can be displayed on a calibration certificate:

UV-Visible spectrophotometers. wavelength

Diffused light

As we have explained previously, the diffused light It is the radiation that can reach the detector with an unwanted wavelength. This radiation can cause significant absorbance reading errors.

The measurement of stray light is performed at wavelengths where the absorbance of the reference standard is greater than 2. Thus, if the measured absorbance were less than 2, it would mean that light is reaching the detector that does not come from the spectrophotometer beam passing through the standard.

Below are data related to the measurement of diffuse light that have been obtained during the calibration of a UV-visible spectrophotometer:

UV-Visible spectrophotometers. scattered light

Absorbance/Photometric Scale

Absorbance is a quantitative measurement that reflects the ability of a substance to absorb light at a specific wavelength. This measurement is crucial in many scientific and technical applications, such as analyzing concentrations of solutions, determining the purity of substances, and studying chemical and biochemical reactions.

Now, it is important to note that, as in the case of wavelength, in a calibration certificate, the data related to absorbance provide information about the correction of the team. As we have explained, the correction is obtained from the subtraction between the certified value and the measured value, with its associated uncertainty:

correction formula

Specifically, the absorbance can be calibrated in the range of wavelengths that each device measures with reference standards of higher and lower absorbances. Thus, in the same way as for the wavelength, the certificate presents in the results the values ​​measured by the equipment to be calibrated, as well as the certified absorbance values ​​associated with the reference standards in each of the measured wavelengths. With these values, the discrepancy between the value measured by the equipment to be calibrated and the certified value of the standard, which appears in the “Correction” column, is calculated.

Below are the data reflected in a calibration certificate for the case of absorbance in the visible range:

UV-Visible spectrophotometers. visible absorbance

Photometric linearity

La photometric scale linearity refers to the linearity of the measured absorbance values ​​as a function of concentration.

To evaluate this linearity, correlation coefficients (R2) of the calibration curve that relates absorbance to concentration. That is, by adjusting the results, it is established how linear the team's behavior is. If it is linear, the equipment is behaving similarly throughout the analyzed wavelength range.

For the following absorbance values ​​in the UV range, the photometric linearity results are shown below.

absorbance values_ photometric linearity

Acceptance/Rejection of calibration results

It is recommended that the company establish an allowable tolerance based on the specific use of the equipment. For example, if it is critical that the equipment measure with a precision of ±0.1 units for the measurements being made, the criteria should reflect that tolerance. Based on this tolerance, it must be evaluated whether the corrections and uncertainties obtained in the calibration are adequate. For this, different criteria can be applied. A common example is the following: C+U ≤ Allowed tolerance.

To define the acceptance and rejection criteria for equipment calibration results, the following key aspects can be considered, based on good metrological practices and international standards:

  1. Compatibility with equipment specifications.
    • Manufacturer specifications: Compare the correction and uncertainty with the equipment manufacturer's precision and accuracy specifications.
    • Allowable Tolerance: Defines a tolerance range based on the specific application of the equipment. For example, if your equipment must measure with a precision of ±0.5 units, your criteria should reflect this tolerance.
  2. Expanded uncertainty (U).
    • Calculation of expanded uncertainty: The standard uncertainty multiplied by a coverage factor (k), generally 2 for a 95% confidence level.
    • Comparison to tolerance: The expanded uncertainty should be small compared to the equipment tolerance. If U is greater than the allowable tolerance, the equipment is not reliable for the required measurement.
  3. Systematic errors.
    • Correction applied: Analyze whether the correction can be applied consistently and accurately in future measurements. If the correction is too large, it could indicate that the equipment needs maintenance or recalibration.
  4. Stability and consistency.
    • Calibration history: Review results from previous calibrations to observe trends. A significant change could indicate a problem with the equipment.
    • Repeatability: Make sure measurements are repeatable under the same conditions.
  5. Impact on the process.
    • Process requirements: Consider how a correction or uncertainty will affect your process or final product. If an incorrect measurement could have serious consequences, the criteria should be more stringent.

Example of acceptance and rejection criteria:

  1. Correction (C) within allowable range:
    1. Accept: Yes |C| ≤ Allowable tolerance
    1. Reject: Yes |C| > Allowable tolerance
  2. Expanded uncertainty (U) within an acceptable limit:
    1. Accept: Yes U ≤ Allowable tolerance/2 (or other factor that is defined)
    1. Reject: Yes U > Allowable tolerance/2
  3. Correction and its uncertainty within the limits of acceptability:
    1. Accept: Yes |C| + U ≤ Allowable tolerance
    1. Reject: Yes |C| + U > Allowable tolerance

Documentation

It is essential to document all calibration criteria and results. The traceability and detailed documentation are essential both for audits and to maintain process quality.

Implementación

To carry out the implementation, a standard operating procedure (SOP) can be developed for the evaluation of the results of the calibration, which includes these criteria and a clear decision flow.

Application of the Lambert-Beer Law

Although the Lambert-Beer Law is not applied to obtain the data that appears in the calibration certificates, we consider it relevant to refer to it, since on a practical level, different industries and laboratories use it as an essential tool to carry out analyzes precise quantitative.

Thus, this law establishes that the absorbance of a substance or species is directly proportional to its concentration. That is, the absorbance of a species in homogeneous solution is directly proportional to its optical activity, optical path length and its concentration. This is an empirical relationship that relates light absorption to the properties of the material passed through.

Below is the absorption diagram of a beam of light passing through a cuvette of length I.

Absorption diagram of a beam of light passing through a bucket of length l
Image 3. Absorption diagram of a beam of light passing through a bucket of length l
legend

The calibration curve corresponds to the Lambert-Beer law and, from it, the concentration of the substances can be obtained. By using solutions with known concentrations, a calibration curve can be created that relates absorbances and concentrations. This can be used as a reference to subsequently be able to obtain the concentration values ​​of the analyte from the absorbance measurements of the spectrophotometer.

Once the calibration line is established, any sample whose concentration is unknown can be analyzed using this relationship. Thus, any absorbance value of unknown concentration interpolated on this calibration line will give us its concentration value. This facilitates the quantification of substances in solution quickly and reliably.

Some concepts

Below, we show some concepts that, beyond those explained above, may be useful.

Difference between traceable calibration and accredited calibration

In traceable calibration, measurements can be traced back to recognized standards, thus ensuring the reliability of the results. It is carried out through a laboratory that uses reference standards or equipment, providing calibration certificates that guarantee this. traceability. The ISO 9001 standard It stipulates that measurements must be traceable to recognized standards, whether national or international, through a continuous chain of comparisons. This traceability guarantees the reliability and validity of the measurements taken.

Furthermore, the accredited calibration It is carried out by laboratories evaluated and accredited by competent entities such as ENACThese laboratories comply with rigorous standards and issue certificates with the ENAC mark, ensuring both traceability and the quality of measurements.

Zero adjustment, verification and recalibration

Zero adjustment is a crucial step before taking measurements with a piece of equipment, where its reading is reset to zero using an adjustment piece supplied by the manufacturer. It is essential to keep this part in perfect condition and understand that it is not interchangeable between equipment. The adjustment ensures the equipment operates according to the manufacturer's specifications.

La check, different from calibration, confirms that the equipment meets the established specifications. It does not involve an adjustment process, but rather the comparison of results with theoretical expectations.

After calibration, if a systematic error is detected, an adjustment can be made to correct it and, subsequently, a recalibration of the equipment. However, not all instruments allow adjustments, nor can all errors be corrected in this way.

Calibration plan

La industry depends largely on a wide range of measuring equipment to ensure the quality of the products it produces. However, to ensure the reliability of these measurements, it is essential that the equipment be subjected to calibration periodically. This process not only ensures that the measurements are accurate, but also ensures that they are traceableThat is, they must be compliant with national or international standards. Therefore, implementing a calibration plan becomes fundamental within a company's quality framework.

The primary objective of the calibration plan is to detect potential errors or malfunctions in the equipment used in the production process. This involves defining the necessary calibrations and determining whether they will be performed internally or externally. In the case of internal calibrations, clear procedures must be established, and it must be ensured that the responsible technical personnel have the appropriate training. It is recommended that the calibration plan also include considerations regarding the acquisition of new instruments or standards, if necessary.

The calibration plan must methodically identify all equipment, standards or measuring instruments, as well as the appropriate calibration intervals for each. This approach ensures that measurements made with these devices maintain the chain of traceability, guaranteeing the integrity of the results.

Additionally, regardless of whether calibrations are performed internally or externally, it is recommended that the company perform an internal check between scheduled calibrations. This additional control helps detect any deviations in measurements and allows corrective action to be taken in a timely manner.

In the case of external calibrations, the company must select accredited laboratories that fit your criteria and needs. These laboratories, with a periodicity established by the company, will carry out the calibration of the equipment and issue certificates with the results obtained. It is essential that the company documents this selection in its calibration plan, thus ensuring the transparency and quality of the process.

Eyco

Eyco offers accredited calibration services and advice on the development of calibration plans to ensure accuracy and reliability in measurements.

Eyco has ENAC accreditation ENAC for the calibration of optical equipment. This ENAC accreditation guarantees the traceability of results and is internationally recognized. Reports and certificates issued under this accreditation are accepted by signatories to EA agreements. ILAC e IAF, which further reinforces the trust that companies and laboratories around the world place in accredited laboratories.

FAQ

What is a UV-Visible spectrophotometer calibration certificate?

Un calibration certificate It is a document issued by a laboratory that indicates the accuracy and reliability of the UV-Visible spectrophotometer. It includes parameters such as error (correction), uncertainty, wavelength accuracy, stray light, and photometric linearity.

Why is it important to interpret calibration certificates correctly?

Interpreting certificates correctly allows verify if the equipment meets the tolerances required and ensure reliable measurements. This is key to ensuring the quality of the tests and the validity of the results in the laboratory.

What does a correction or error in a certificate mean?

La correction or error This represents the difference between the value measured by the spectrophotometer and the reference value. It allows for adjusting the results and evaluating the accuracy of the equipment.

What is measurement uncertainty?

La uncertainty It indicates the margin of doubt associated with a measurement. In certificates, it is usually shown as expanded uncertainty with a 95% confidence level, helping to determine the reliability of the results.

What is the difference between traceable and accredited calibration?

La traceable calibration is linked to internationally recognized standards, while the accredited calibration It is carried out in laboratories accredited by organizations such as ENAC, guaranteeing the validity and official recognition of the results.

Why is a calibration plan important in the laboratory?

Un calibration plan Define which equipment to calibrate, how often and under what criteria, ensuring traceability, quality of measurements and reliability of results for analysis and quality control.