Contact us Menu

Ensuring the world is accurate

FTIR

FTIR Spectrophotometer Calibrations

On-site or laboratory calibration, with a personalized and flexible service. Delivery time: 5 days.


Calibration consists of establishing, under specific conditions, the relationship between the values ​​measured by a piece of equipment and the reference values ​​traceable to recognized standards.

This relationship allows the client to evaluate the instrument's performance and ensure the validity of its measurements. Calibration focuses on FTIR (Fourier Transform Infrared) spectrophotometers, which are based on the interaction of infrared radiation with matter. Specifically, these instruments are crucial for identifying chemical substances, functional groups, and the molecular structure of organic materials, plastics, or paints, guaranteeing the metrological traceability of results in the mid-infrared region (4000 cm⁻¹).-1 to 400cm-1.

Wavenumber scale calibration (cm−1)

FTIR spectrophotometry provides a spectrum where absorbance/transmittance is related to wavenumber (cm−1Specifically, calibration consists of verifying the accuracy of this scale, which is fundamental for accurately determining the location of the peaks or spectral bands (e.g., C=O at 1700 cm).−1), which are the "fingerprint" for identifying the chemical substance analyzed.

Specific techniques (ATR)

We calibrate FTIR equipment that uses common techniques such as Attenuated Total Reflection (ATR), which allows the analysis of thick and opaque samples (viscous liquids, plastics, coatings) without the need for complex preparation.

What equipment do we calibrate?

FTIR Spectrophotometers

On the one hand, the calibration of FTIR spectrophotometers designed to work in the mid-infrared region. Specifically, we focus on the calibration of the wavenumber scale (cm−1), which is the key parameter for the structural identification of functional groups in solid and liquid samples.

Calibration Description For this type of equipment, Eyco performs accuracy calibration (the number of points to be calibrated will depend on the range of the equipment):

  • Calibration of wave number scale accuracy at nominal points: 539, 841, 906, 1028, 1069, 1154, 1582, 1601, 1942, 2849, 3001, 3025, 3060, 3082 cm-1.

Determination of the repeatability of the equipment.

Additionally, you may request a determination of the decision-making power (outside the scope of ENAC), indicating it in the comments.

Polystyrene Films or Sheets

Furthermore, we also calibrate reference materials such as polystyrene films, which are essential for verifying the wavenumber scale of FTIR spectrophotometers. Specifically, polystyrene exhibits very distinct and well-known absorption bands. By calibrating this material, we ensure that its reference values ​​are accurate. This allows the client to perform periodic checks, intercomparisons between equipment, or internal calibrations with complete reliability.

Calibration of FTIR Spectrophotometers

Why Is This Important?

Regular calibration of FTIR spectrophotometers is essential for unambiguous chemical identification. This directly impacts quality control, research, and product safety.

Technical scope for FTIR calibration

We can calibrate equipment that measures by absorbance/transmittance or with the ATR accessory.

Measurement range
Range
Uncertainty (*)
--
Standard/Procedure
--
Instruments to be calibrated
Instruments
Dress
Code
Wave number
Wave number
500 cm-1 ≤λ-1 ≤ 800 cm-1
800 cm-1 ≤λ-1≤ 3100 cm-1
1,6 cm-1
1,0 cm-1
PT38
Procedure
internal based on:
Royal Pharmacopoeia
Spanish
FTIR SpectrophotometersA,I

Technical scope for polystyrene calibration

Measurement range
Range
Uncertainty (*)
--
Standard/Procedure
--
Instruments to be calibrated
Instruments
Dress
Code
Wave number
Wave number
400 cm-1 ≤ 1/λ ≤ 4000cm-11.7PT51Polystyrene filmsA

The result

After the equipment is calibrated, you will receive the Calibration Certificate backed by accreditation. ENACSpecifically, this Certificate includes the complete identification of the equipment, the procedure followed for calibration, and the result.

The calibration result is the correction based on the wavenumber and its expanded uncertainty. The correction is the difference between the reference standard value and the value measured by your instrument. Based on the correction values ​​and their uncertainty, you can then establish the instrument's acceptance and rejection criteria.

If a polystyrene film needs to be calibrated, the calibration certificate will provide the wavenumber values ​​for which the spectral bands are found. Based on these values ​​and their uncertainty, the acceptance and rejection criteria for the filters can then be established.

Also, together with the equipment or material, you will receive a calibration label with the certificate number, equipment identification and calibration date.

More about FTIR spectrophotometer calibrations

Frequently Asked Questions. FTIR Spectrophotometer Calibrations