Focimeter calibrations
Calibration in the laboratory or on-site, with a personalized and flexible service. Delivery time: 5 days.
What do we calibrate?
ENAC accredited frontofocometer calibration service.
Calibration of measurement accuracy at 10 points, distributed between +25 and −25 diopters, provided that the range of the equipment allows it.
Eyco is the only laboratory in Spain accredited by ENAC for the calibration of lensometers.
Focimeters
The standard instrument used to measure optical power is the lensometer. Specifically, this device is designed to measure the front and back power of ophthalmic lenses.
Therefore, the accuracy and quality of the measurements taken with this equipment is crucial, since it is essential to ensure that the lenses have the required power. Like any measuring instrument, lens gauges have an associated measurement uncertainty that can be determined through calibration.
Calibrating these devices ensures the accuracy of the optical power measurements provided by the lensometer. Furthermore, regular calibration allows for the timely detection of errors or drift in the equipment due to usage, component aging, and other factors.

Technical scope for the calibration of focimeters
| Measurement range Range | Uncertainty (*) -- | Standard/Procedure -- | Instruments to be calibrated Instruments | Dress Code |
|---|---|---|---|---|
| OPTICAL POWER OPTICAL POWER | ||||
| -25 ≤ PO ≤ +25 diopters (m-1) | 0,020 diopters (m-1) | PT23 Internal procedure based on: UNE-EN ISO 9342-1 UNE-EN ISO 8598-1 | Focimeters | A,I |
The result
Once your equipment has been calibrated by Eyco's expert staff, you will receive a Calibration Certificate backed by accreditation. ENAC. Said Certificate includes the complete identification of your equipment, the procedure that the laboratory has followed to carry out the calibration and the result of the same.
The result of lens calibration is a correction along with its uncertainty. The correction is the difference between the reference standard value and the value measured by your instrument.
From the correction values and their uncertainty, you can establish the acceptance and rejection values of the equipment.
In addition, together with your equipment, you will receive a calibration label with the certificate number, the identification of the equipment and the date of calibration.