Spectrophotometer and colorimeter calibrations

On-site or laboratory calibration, with a personalized and flexible service. Delivery time: 5 days.

Contact

What do we calibrate?

Calibration of spectrophotometers and colorimeters.

Calibration can be performed for one or more measurement conditions.

Each calibration condition It is defined as a specific combination of:

  • a measurement geometry (0º:45º, 8º:di or 8º:de),
  • un illuminant (A, C, D65 or D50),
  • or with a observer (2nd or 10th).

Determination of the repeatability of the equipment on a white pattern.

ENAC accredited calibration.

Color evaluation

In many industrial sectors, color is an important variable to control. Many companies are concerned with evaluating the color of their products and establishing color tolerances.
To do this, it is essential to objectively measure color, using instruments such as colorimeters, spectrophotometers or spectrocolorimeters.
They ensure quality control of products in a wide variety of industrial sectors, such as the production of foods, paintingsceramics, toys, textilescosmetics, plastic, chemical sectorTherefore, it is essential to keep these instruments properly calibrated. accredited laboratory.

Calibration of spectrophotometers and colorimeters

How do colorimeters, spectrophotometers, or spectrocolorimeters measure color?

The most common way to objectively measure color is to take L*a*b* coordinate measurements and calculate the color differences between a standard and a sample, such as DE*. However, some sectors, such as signage, use Yxy coordinates for color evaluation.

At Eyco we adapt to your company's needs and perform the calibration you require according to your industry (textile, painting, eating, chemistry…

Color measurement conditions

Color is not an intrinsic property of materials, but is the result of the interaction of radiation (light/illuminant) with an object, detected by a detector (observer) and taking into account the way in which light interacts with the object (geometry).

To measure color, it is necessary to know the measurement conditions: geometry, illuminant, and observer. Any change in the measurement conditions will result in changes in the color coordinates. Therefore, it is very important that the company requesting the calibration provides the laboratory with the measurement conditions it normally uses.

Calibration of spectrophotometers and colorimeters
Geometry

Geometry 45º:0º
 With this geometry, the instrument will illuminate the sample from a 45° angle and measure the light reflected at 0°; whereas if the instrument geometry is 0°:45°, it will illuminate the sample at 0° and detect the light reflected at 45°. This geometry is commonly used in the graphic arts sector and in the signage industry, since it excludes glare from the color measurement.

Sphere geometry
 Integrated Illumination Sphere systems can perform measurements in both Specular Mode Included (SCI) and Specular Mode Excluded (SCE). The sphere geometry is the most widespread in all sectors.

Geometry 0º:0º
 This measurement method is used in equipment that measures color by transmission, for both liquids and translucent solids (glass, plastics, etc.). Eyco is the only laboratory in Spain accredited for calibrating equipment with this geometry.

Illuminant

The illuminant is the type of light that falls on the sample that we want to measure the color. The illuminant is a spectral distribution curve of light sources.

The most common illuminants in colorimeters and spectrophotometers are: D65, D50, A and C. Specifically, illuminants D65 and D50 correspond to daylight. The first is widely used in a variety of sectors, while the second is usually common in the graphic arts sector. As for illuminant A, it refers to artificial light, specifically, it simulates incandescent light.

In addition, there are other illuminants that simulate fluorescent light sources, although their use is usually limited to very specific applications, or to certain test regulations that require it.

Observer

Without an observer, color cannot be detected and therefore measured. The types of observers are standardized by the CIE: 2º and 10º. Each of them represents a different visual field. The 10º observer is the most common, although it always depends on the type of company and sector.

Technical scope for the calibration of spectrophotometers and colorimeters

Measurement range
Range
Uncertainty (*)
--
Standard/Procedure
--
Instruments to be calibrated
Instruments
Dress
Code
Color, LabY xy Coordinates Color difference dE (8°:di and 8°:de)
Color, LabY x coordinates,and Color difference dE (8°:di and 8°:de)
0 ≤ L*≤ 100

– 100 ≤ a*≤ 100

– 100 ≤ b*≤ 100

0 ≤ Y ≤ 100

0 ≤ x ≤ 0,9

0 ≤ y ≤ 0,9

0 ≤ dE*≤ 10
0,30

0,20

0,30

0,40

0,0020

0,0020

0,30
PT40
Internal procedure based on: IUPAC
Recommended Reference Materials for realization of Physicochemical properties CIE 15
Spectrophotometers
Tristimulus colorimeters
A,I
Color, LabY xy Coordinates Color difference dE (0°:45a)
Color, LabY x coordinates,and Color difference dE (0º:45a)
0 ≤ L*≤ 100

– 100 ≤ a*≤ 100

– 100 ≤ b*≤ 100

0 ≤ Y ≤ 100

0 ≤ x ≤ 0,9

0 ≤ y ≤ 0,9

0 ≤ dE*≤ 10
0,30

0,20

0,30

0,40

0,0020 0,0020

0,30
PT40 Internal procedure based on: IUPAC Recommended Reference Materials for realization of Physicochemical properties CIE 15Spectrophotometers
Tristimulus colorimeters
Spectrodensitometers
AI

The result

Once your equipment has been calibrated by Eyco's expert staff, you will receive a Calibration Certificate backed by accreditation. ENAC. Said Certificate includes the complete identification of your equipment, the procedure that the laboratory has followed to carry out the calibration and the result of the same.

The result of spectrophotometer and colorimeter calibration is a correction along with its uncertainty. This correction is the difference between the value of the reference standard and the value measured by your instrument. The Calibration Certificate issued by Eyco will provide the chromaticity coordinate correction for six standards and the color difference correction.


From the correction values ​​and their uncertainty, you can establish the acceptance and rejection values ​​of the equipment.

In addition, together with your equipment, you will receive a calibration label with the certificate number, the identification of the equipment and the date of calibration.