Standard Lens Calibrations
Laboratory calibration, with a personalized and flexible service. Delivery time: 5 days.
What do we calibrate?
Eyco is the only laboratory in Spain accredited by ENAC for calibration of standard lenses.

Standard lens calibration
A set of standard lenses consists of calibrated lenses used for calibrating and verifying lensometers. The characteristics that these lenses must meet are detailed in Standard 9342. Specifically, point 4 of the Standard indicates that they must have the following characteristics:
- On the one hand, they must not have striations in an area of 4 mm radius around the center of the free opening.
- Furthermore, the standard lenses must be manufactured from homogeneous Crown glass with a refractive index nd=1.523±0.002 or ne= 1.525 ± 0.002.
- In addition, the reference wavelength for which the standard lenses are calibrated must be specified.
- Finally, the lenses should have a protective frame, designed so that when the lenses are correctly placed in the holder, the lens gauge is not obstructed.
Technical scope for standard lens calibration
| Measurement range Range | Uncertainty (*) -- | Standard/Procedure -- | Instruments to be calibrated Instruments | Dress Code |
|---|---|---|---|---|
| OPTICAL POWER OPTICAL POWER | ||||
| -25 ≤ PO ≤ +25 diopters (m-1) | 0,040 diopters (m-1) | PT32 Internal procedure based on: UNE-EN ISO 9342-1 UNE-EN ISO 8598-1 | Spherical ophthalmic lenses pattern | A |
The result
After the lenses have been calibrated, you will receive a Calibration Certificate backed by accreditation. ENAC. Said Certificate includes the complete identification, the procedure that the laboratory has followed to carry out the calibration and the result of the same.
The result of the calibration is the value of the optical power in Diopters along with its uncertainty, for each of the standard lenses.
From the correction values and their uncertainty, you can establish the acceptance and rejection values of the lenses.
Also, along with your lenses, you will receive a calibration label with the certificate number, identification and date of calibration.