Contact us Menu

Ensuring the world is accurate

IMG_9637

Color Standard Calibrations

Personalized and flexible service. Delivery time: 5 days.

Contact

What do we calibrate?

Calibration of color patterns for a measurement condition: 1 geometry, 1 observer and 1 illuminant.

Color Standards

Color patterns are used for daily equipment verification, intercomparison of various equipment within the same company, periodic checks, or even internal equipment calibrations.

Color standards are calibrated for the necessary geometry, illuminant, and observer conditions.

Geometry

Geometry 45º:0ºThis geometry is commonly used in the sector of graphic arts and in the signage industry, as it excludes brightness in the measurement of color.

Geometry of a sphere. Integrated Illumination Sphere systems can perform Specular Mode Included (SCI) and Specular Mode Excluded (SCE) measurements.

Sphere geometry is the most widespread in all sectors.

Observer

Without an observer, color cannot be detected or measured. The CIE standardizes two types of observers: 2° and 10°. Each represents a different field of vision. Specifically, the 10° observer is the most common, although this always depends on the type of company and industry.

Illuminant

The illuminant is the type of light that falls on the sample whose color we want to measure. The illuminant is a spectral distribution curve of light sources.

The most common illuminants in Colorimeters and Spectrophotometers These are: D65, D50, A, and C. Specifically, illuminants D65 and D50 correspond to daylight. The former is widely used in a variety of sectors, while the latter is common in the graphic arts industry. As for illuminant A, it refers to artificial light (it simulates incandescent light).

There are also illuminants that reproduce fluorescent light sources, although their application is restricted to specific cases or to the requirements established by certain testing standards.

Color Standard Calibration

Standards for colorimeters and spectrophotometers

The  Colorimeters and Spectrophotometers They are usually supplied with a reference standard. This standard is typically ceramic and white, although a green standard is sometimes also provided. These standards are characterized by their stability and require some maintenance. Specifically, they are usually used for the routine (usually daily) adjustment of the equipment, so that it is possible to verify an accurate color measurement.

It is recommended to properly maintain these standards and avoid dust, fingerprints, damage, scratches, etc. on them. To verify that its value in chromatic coordinates is not altered by time, periodic verifications can be carried out. Color standards are calibrated under the same geometry, illuminant, and observer conditions in which the equipment is used, and can be calibrated in both CIELab* and Yxy coordinates.

When color standards are used for equipment intercomparison within the same company, standards must be available that have been maintained in good condition and have been calibrated, where the values ​​of the chromatic coordinates are known and stable, so that the same standard can be measured by different equipment.

Thus, for example, equipment that is used in the laboratory could be compared in the same company with equipment that is used in the production plant and evaluate if there is any deviation that could lead to rejection of batches of products.

Color standards can also be used by a company for periodic checks or during the periods between external calibrations of the colorimeter or spectrophotometer. This ensures that the equipment is measuring correctly and eliminates any doubts about potentially inaccurate measurements or the accuracy of the measurement conditions. Having a calibrated color standard guarantees the accuracy of its chromaticity coordinates.

Some companies choose to have a complete set of standards that allow them to perform their own internal calibrations. In these cases, the correct approach is to use a set of standards that cover the entire color diagram and have an internal calibration procedure, along with an uncertainty calculation.

Color Standards

Technical scope for calibration of color standards

Measurement range
Range
Uncertainty (*)
--
Standard/Procedure
--
Instruments to be calibrated
Instruments
Dress
Code
Color, coordinates L*a*b*Y xy Color difference dE* (8°:di and 8°:de)
Color, coordinates L*a*b*Y xy, Color difference dE* (8°:di and 8°:de)
0 ≤ L*≤ 100
– 100 ≤ a*≤ 100
– 100 ≤ b*≤ 100
0 ≤ Y ≤ 100
0 ≤ x ≤ 0,9
0 ≤ y ≤ 0,9
0,50
0,50
0,50
0,50
0,0020
0,0020
PT45
Procedure
internal based on:
IUPAC Recommended
Reference Materials
for realization of
Physicochemical
properties
Color StandardsA
Color, Coordinates L*a*b*Y xy Color difference dE* (0°:45a)
Color, coordinates L*a*b*Y xy, Color difference dE* (0°:45a)
0 ≤ L*≤ 100
– 100 ≤ a*≤ 100
– 100 ≤ b*≤ 100
0 ≤ Y ≤ 100
0 ≤ x ≤ 0,9
0 ≤ y ≤ 0,9
0,50
0,50
0,50
0,50
0,0020
0,0020
PT45
Procedure
internal based on:
IUPAC Recommended
Reference Materials
for realization of
Physicochemical
properties
Color StandardsA

The result

Once your equipment has been calibrated by Eyco's expert staff, you will receive a Calibration Certificate backed by accreditation. ENAC. Said Certificate includes the complete identification of your equipment, the procedure that the laboratory has followed to carry out the calibration and the result of the same.

The result of color pattern calibration is the value of the color units of each pattern for each geometry along with its expanded uncertainty.

From the correction values ​​and their uncertainty, you can establish the acceptance and rejection criteria.

Also, along with your standard, you will receive a calibration label with the certificate number, standard identification and calibration date.

Frequently Asked Questions. Color Pattern Calibration