Filter calibrations
Laboratory calibration, with a personalized and flexible service. Delivery time: 5 days.
What do we calibrate?
- Filters for wavelength calibration
- Filters for absorbance and transmittance calibration
- Filters for linearity calibration
- Filters for stray light calibration
Filters or standards
Filters or standards can have different uses within a company, from daily verification of a UV-Visible spectrophotometer, intercomparison of various equipment within the same company, periodic checks or even to perform internal calibrations.
Thus, depending on the use of the equipment, the company can have filters for calibrating the wavelength, optical density, transmittance, linearity of the spectrophotometer and diffuse light.
Specifically, these filters can be solid or liquid. Solid filters are glass mounted in 1cm spectrophotometer cuvettes. Liquid filters, on the other hand, come in sealed 1cm spectrophotometer cuvettes.

Filters for wavelength calibration
The filters used to calibrate the wavelength scale of spectrophotometers are filters with well-defined, narrow absorption bands for different wavelengths. Specifically, the usual standards for these calibrations are holmium oxide and didymium oxide filters.
On the one hand, the holmium oxide filter exhibits absorbance peaks in the UV-Visible range, from 240nm to 650nm.
Furthermore, the didymium filter also exhibits absorbance peaks in the UV-Visible range, from 380nm to 900nm.
Specifically, Eyco has accreditation ENAC for the calibration of holmium and didymium oxide filters and any other filter used for wavelength calibration in the UV-Visible range.
We can also calibrate filters for the wavelength in the infrared, from 800nm to 1800nm.
| Measurement range Range | Uncertainty (*) -- | Standard/Procedure -- | Instruments to be calibrated Instruments | Dress Code |
|---|---|---|---|---|
| WAVELENGTH Waveength | ||||
| 200nm ≤ λ ≤ 850nm 850nm ≤ λ ≤ 2000nm | 0.30 nm 0.50 nm | PT43 Internal procedure based on ICD 130 | Wavelength filters | A |
Filters for absorbance and transmittance calibration
Filters or standards for the calibration of the photometric scale are used to determine the accuracy of the absorbance and transmittance values measured by the spectrophotometer.
These standards or filters are typically neutral density glass mounted on supports compatible with 1 cm spectrophotometer cuvettes. Specifically, these standards display different absorbance or transmittance values at various wavelengths. Therefore, combining or using several filters allows for determining the instrument's error or correction at different points on the photometric or absorbance and transmittance scales.
Furthermore, for the ultraviolet region, potassium dichromate solutions are also very commonly used. Specifically, these reference materials have established absorbance values at different wavelengths.
Specifically, Eyco has accreditation ENAC for the calibration of filters or standards in the ultraviolet-visible range, for different absorbance values.
We can also calibrate filters in the infrared range from 780nm to 2500nm.
| Measurement range Range | Uncertainty (*) -- | Standard/Procedure -- | Instruments to be calibrated Instruments | Dress Code |
|---|---|---|---|---|
| TRANSMITTANCE TRANSMITTANCE | ||||
| 200 nm ≤ λ ≤ 420 nm 0 ≤ T < 0,5 0,5 ≤ T < 1,0 1,0 ≤ T < 2,0 2,0 ≤ T ≤ 2,5 | 0,0045·T 0,0065·T 0,025·T 0,065·T | PT43 Internal procedure based on ICD 130 | Transmittance filters | A |
| (420-900)nm 0.00010≤ T <0.0010 0.0010≤T<0.10 0.10≤ T ≤1 | 0.35T 0.015T 0.0070T | PT43 Internal procedure based on ICD 130 | Transmittance filters | A |
| (900-2500)nm 0.010≤ T <0.20 0.20≤ T ≤1 | 0.025T 0.0070T | PT43 Internal procedure based on ICD 130 | Transmittance filters | A |
| TRANSMITTING OPTICAL DENSITY (ABSORBANCE) OPTICAL TRANSMITTANCE DENSITY (ABSORBANCE) | ||||
| 200 nm ≤ λ ≤ 420 nm 0 ≤ A < 0,5 0,5 ≤ A < 1,0 1,0 ≤ A < 2,0 2,0 ≤ A ≤ 2,5 | 0,0045 0,0065 0,025 0,065 | PT43 Internal procedure based on ICD 130 | Absorbance filters | A |
| (420-900)nm 0≤A<1.50 1.50≤A≤3 3≤A≤4 | 0.0045 0.055 0.25 | PT43 Internal procedure based on ICD 130 | Absorbance filters | A |
| (900-2500)nm 0≤A≤1 1≤A≤2 | 0.0045 0.025 | PT43 Internal procedure based on ICD 130 | Absorbance filters | A |
Filters for linearity calibration
Furthermore, linearity calibration filters are used to assess the accuracy of spectrophotometers in determining the absorbance of solutions as a function of concentration. For this purpose, different standards with varying concentrations are available. Additionally, potassium dichromate solutions are commonly used as reference material, and the instrument's response is evaluated in terms of absorbance or transmittance.
Specifically, at Eyco we can perform the calibration of these filters in absorbance and transmittance within the accreditation ENAC.
Filters for stray light calibration
Finally, these filters or patterns are useful for assessing whether the detector is receiving light of a different wavelength than the required one. Specifically, the instrument's detector cannot distinguish between light passing through the sample and light reaching the instrument from other sources.
Filters for evaluating stray light are typically solutions in sealed 1cm cuvettes for spectrophotometers. Specifically, these filters absorb all light above a certain wavelength. Therefore, if the detector detects light, it means the light source is not the sample and is therefore stray light.
These filters are usually solutions of NaI, KCl, etc.
The result
After the calibration of the standards, you will receive the Calibration Certificate covered under the accreditation. ENACSpecifically, this Certificate includes the full identification of your equipment, the calibration procedure followed, and the result.
First, for wavelength filters, the calibration result is the value of the absorbance peaks of each standard along with its expanded uncertainty.
On the other hand, for absorbance or transmittance filters, the calibration result is the absorbance value of each standard along with its expanded uncertainty.
Based on the values of the certificates and the uncertainty, the company can establish the acceptance and rejection criteria.
In addition, along with the pattern, you will also receive a calibration label with the certificate number, equipment identification, and calibration date.