Microplate calibrations

On-site or laboratory calibration, with a personalized and flexible service. Delivery time: 5 days.

Contact

What do we calibrate?

Microplate calibration.

Depending on the characteristics of the microplate, calibration is carried out as follows:

  • Wavelength calibration: Determination of absorbance peaks, if the microplate allows it. Maximum 5 points (279 nm, 361 nm, 453 nm, 536 nm and 637 nm)
  • Absorbance calibration: Determination of absorbance within the range of 200nm to 2500nm, at a maximum of 4 points (If the points are not specified, the laboratory will calibrate the equipment at: 405nm, 450nm, 490nm, and 650nm). (For additional calibration points, please consult)

Microplates

Microplates are used for internal verification or calibration of microplate readers or ELISA spectrophotometers. With them it is possible to calibrate the wavelength scale and the photometric scale of the equipment.

Specifically, microplates, also called standard plates, are usually rectangular, with approximate dimensions of 12.5 cm by 8.5 cm, and consist of 8 rows and 12 columns containing 96 wells. Each of these 12 columns represents a specific absorbance value at different wavelengths. In addition, in some cases, there are columns used for wavelength scale calibration, which contain a holmium oxide filter with known absorbance peaks.

Calibration in absorbance

On the one hand, the calibration in absorbance, or the calibration of the photometric scale, allows measuring the accuracy or correction that the equipment presents with respect to the absorbance values ​​for different wavelengths.

Specifically, the standard plates consist of 5 or 6 neutral density filters with different absorption values, arranged along the columns. The absorbance values ​​of these filters range from 0.2 OD to 3.5 OD.

Wavelength Calibration

On the other hand, some microplate models include columns for wavelength scale calibration. This type of calibration allows for the evaluation of the spectrophotometer's monochromator component's ability to isolate a specific wavelength from the rest of the light spectrum.

Specifically, these microplates, which have this characteristic, contain filters in their columns with narrow absorption bands for different wavelengths. These filters are usually made of holmium oxide and exhibit absorbance peaks in the UV-visible range, from 240 nm to 650 nm.

Microplate calibration

Technical scope for microplate calibration

Eyco has accreditation ENAC for calibration of wavelength filters and neutral density filters.

Measurement range
Range
Uncertainty (*)
--
Standard/Procedure
--
Instruments to be calibrated
Instruments
Dress
Code
TRANSMITTANCE
TRANSMITTANCE
200 nm ≤ λ ≤ 420 nm
0 ≤ T < 0,5
0,5 ≤ T < 1,0
1,0 ≤ T < 2,0
2,0 ≤ T ≤ 2,5
0,0045·T
0,0065·T
0,025·T
0,065·T
PT43 Internal procedure based on ICD 130F Absorbance filters
Density filters
neutral
Optical filters
Microplates
A
(420-900)nm
0.00010≤T<0.0010
0.0010≤T<0.10
0.10≤ T ≤1
0.35T
0.015T
0.0070T
PT43 Internal procedure based on ICD 130Transmittance filters
Optical filters
Microplates
A
(900-2500)nm
0.010≤T<0.20
0.20≤ T ≤1
0.025T
0.0070T
PT43 Internal procedure based on ICD 130Transmittance filters
Optical filters
Microplates
A
TRANSMITTING OPTICAL DENSITY (ABSORBANCE)
OPTICAL TRANSMITTANCE DENSITY (ABSORBANCE)
200 nm ≤ λ ≤ 420 nm
0 ≤ A < 0,5
0,5 ≤ A < 1,0
1,0 ≤ A < 2,0
2,0 ≤ A ≤ 2,5
0,0045
0,0065
0,025
0,065
PT43 Internal procedure based on ICD 130Absorbance filters Neutral density filters
Optical filters
Microplates
A
(420 – 900)nm
0 ≤ A ≤ 1,50
1,50 ≤ A ≤ 3
3 ≤ A ≤ 4
0,0045
0,055
0,25
PT43 Internal procedure based on ICD 130Absorbance filters Neutral density filters
Optical filters
Microplates
A
(900-2500)nm
0≤A≤1
1≤A≤2
0.0045
0.025
PT43 Internal procedure based on ICD 130Absorbance filters Neutral density filters
Optical filters
Microplates
A
Measurement range
Range
Uncertainty (*)
--
Standard/Procedure
--
Instruments to be calibrated
Instruments
Dress
Code
WAVELENGTH
Waveength
200nm≤λ≤850nm
850nm ≤λ ≤2000nm
0.30nm
0.50nm
PT43 Internal procedure based on ICD 130Microplate FiltersA

The result

Once your equipment has been calibrated by Eyco's expert staff, you will receive a Calibration Certificate backed by accreditation. ENACSpecifically, the Certificate includes the complete identification of your equipment, the procedure the laboratory followed to perform the calibration, and the result of the calibration.

Specifically, the result of calibration in the case of absorbance or transmittance filters is the absorbance value of each standard along with its expanded uncertainty.

Thus, based on the correction values ​​and their uncertainty, you can establish the acceptance and rejection values ​​of the equipment.

In addition, together with your equipment, you will receive a calibration label with the certificate number, the identification of the equipment and the date of calibration.

Frequently Asked Questions. Microplate Calibration