Microplate calibrations
On-site or laboratory calibration, with a personalized and flexible service. Delivery time: 5 days.
What do we calibrate?
Microplate calibration.
Depending on the characteristics of the microplate, calibration is carried out as follows:
- Wavelength calibration: Determination of absorbance peaks, if the microplate allows it. Maximum 5 points (279 nm, 361 nm, 453 nm, 536 nm and 637 nm)
- Absorbance calibration: Determination of absorbance within the range of 200nm to 2500nm, at a maximum of 4 points (If the points are not specified, the laboratory will calibrate the equipment at: 405nm, 450nm, 490nm, and 650nm). (For additional calibration points, please consult)
Microplates
Microplates are used for internal verification or calibration of microplate readers or ELISA spectrophotometers. With them it is possible to calibrate the wavelength scale and the photometric scale of the equipment.
Specifically, microplates, also called standard plates, are usually rectangular, with approximate dimensions of 12.5 cm by 8.5 cm, and consist of 8 rows and 12 columns containing 96 wells. Each of these 12 columns represents a specific absorbance value at different wavelengths. In addition, in some cases, there are columns used for wavelength scale calibration, which contain a holmium oxide filter with known absorbance peaks.
Calibration in absorbance
On the one hand, the calibration in absorbance, or the calibration of the photometric scale, allows measuring the accuracy or correction that the equipment presents with respect to the absorbance values for different wavelengths.
Specifically, the standard plates consist of 5 or 6 neutral density filters with different absorption values, arranged along the columns. The absorbance values of these filters range from 0.2 OD to 3.5 OD.
Wavelength Calibration
On the other hand, some microplate models include columns for wavelength scale calibration. This type of calibration allows for the evaluation of the spectrophotometer's monochromator component's ability to isolate a specific wavelength from the rest of the light spectrum.
Specifically, these microplates, which have this characteristic, contain filters in their columns with narrow absorption bands for different wavelengths. These filters are usually made of holmium oxide and exhibit absorbance peaks in the UV-visible range, from 240 nm to 650 nm.

Technical scope for microplate calibration
Eyco has accreditation ENAC for calibration of wavelength filters and neutral density filters.
| Measurement range Range | Uncertainty (*) -- | Standard/Procedure -- | Instruments to be calibrated Instruments | Dress Code |
|---|---|---|---|---|
| TRANSMITTANCE TRANSMITTANCE | ||||
| 200 nm ≤ λ ≤ 420 nm 0 ≤ T < 0,5 0,5 ≤ T < 1,0 1,0 ≤ T < 2,0 2,0 ≤ T ≤ 2,5 | 0,0045·T 0,0065·T 0,025·T 0,065·T | PT43 Internal procedure based on ICD 130 | F Absorbance filters Density filters neutral Optical filters Microplates | A |
| (420-900)nm 0.00010≤T<0.0010 0.0010≤T<0.10 0.10≤ T ≤1 | 0.35T 0.015T 0.0070T | PT43 Internal procedure based on ICD 130 | Transmittance filters Optical filters Microplates | A |
| (900-2500)nm 0.010≤T<0.20 0.20≤ T ≤1 | 0.025T 0.0070T | PT43 Internal procedure based on ICD 130 | Transmittance filters Optical filters Microplates | A |
| TRANSMITTING OPTICAL DENSITY (ABSORBANCE) OPTICAL TRANSMITTANCE DENSITY (ABSORBANCE) | ||||
| 200 nm ≤ λ ≤ 420 nm 0 ≤ A < 0,5 0,5 ≤ A < 1,0 1,0 ≤ A < 2,0 2,0 ≤ A ≤ 2,5 | 0,0045 0,0065 0,025 0,065 | PT43 Internal procedure based on ICD 130 | Absorbance filters Neutral density filters Optical filters Microplates | A |
| (420 – 900)nm 0 ≤ A ≤ 1,50 1,50 ≤ A ≤ 3 3 ≤ A ≤ 4 | 0,0045 0,055 0,25 | PT43 Internal procedure based on ICD 130 | Absorbance filters Neutral density filters Optical filters Microplates | A |
| (900-2500)nm 0≤A≤1 1≤A≤2 | 0.0045 0.025 | PT43 Internal procedure based on ICD 130 | Absorbance filters Neutral density filters Optical filters Microplates | A |
| Measurement range Range | Uncertainty (*) -- | Standard/Procedure -- | Instruments to be calibrated Instruments | Dress Code |
|---|---|---|---|---|
| WAVELENGTH Waveength | ||||
| 200nm≤λ≤850nm 850nm ≤λ ≤2000nm | 0.30nm 0.50nm | PT43 Internal procedure based on ICD 130 | Microplate Filters | A |
The result
Once your equipment has been calibrated by Eyco's expert staff, you will receive a Calibration Certificate backed by accreditation. ENACSpecifically, the Certificate includes the complete identification of your equipment, the procedure the laboratory followed to perform the calibration, and the result of the calibration.
Specifically, the result of calibration in the case of absorbance or transmittance filters is the absorbance value of each standard along with its expanded uncertainty.
Thus, based on the correction values and their uncertainty, you can establish the acceptance and rejection values of the equipment.
In addition, together with your equipment, you will receive a calibration label with the certificate number, the identification of the equipment and the date of calibration.