Microplate Reader calibrations

On-site or laboratory calibration, with a personalized and flexible service. Delivery time: 5 days.

Contact

What do we calibrate?

Calibration of microplate reader and ELISA spectrophotometer.

  • Determination of wavelength accuracy to 5 points from 250nm to 650nm (if the equipment allows it).
  • Determination of the accuracy of the photometric scale (absorbance) within the instrument's range: a maximum of 4 points within the instrument's range of 400 to 900 nm. If the points are not specified, the laboratory will calibrate the instrument at: 405 nm, 450 nm, 490 nm, and 650 nm, or at any 4 points the instrument allows. (For additional calibration points, please consult.)
  • Determination of repeatability of absorbance readings.

Microplate readers

A microplate reader is an optical device that detects biological, chemical, or physical processes or phenomena by measuring absorbance. Specifically, they are known as ELISA readers or spectrophotometers.Enzyme Linked Immunosorbent Assay-Enzyme-Linked Immunosorbent Assay), associated with a laboratory technique for the determination of antigens in samples.

The physical principle of the instrument is identical to that of a UV-visible spectrophotometer: it measures the amount of light transmitted or absorbed by a sample at different wavelengths. It consists of a light source, a monochromator, and a detector.

Alternative

Lighting system

On the one hand, microplate readers most commonly use a tungsten light source for the visible region, although there are also devices on the market that reach the ultraviolet range. Specifically, unlike a conventional UV-visible spectrophotometer, a microplate reader typically operates between 400 nm and 700 nm.

Detector

On the other hand, the detector is responsible for transforming the signal it receives (photons) and converting it into an electrical signal, so that depending on its intensity it can be recorded by a system that provides the absorbance or DO numerical values.

The calibration of the detector scale consists in determining the correctness of the absorbance values ​​for the given wavelengths and their uncertainty.

Monochrome

Finally, the monochromator allows the separation of the light spectrum by providing a beam of energy at a specific wavelength. This enables work at a particular wavelength. For example, if working with different chemical compounds that exhibit a characteristic absorbance at 690 nm, it is important that the monochromator precisely separates this wavelength from the rest. It would be necessary to ensure that the equipment operates at 690 nm and not 693 nm.

Specifically, if the wavelength is critical or affects the measurements being taken, it is necessary to calibrate this part of the equipment, known as wavelength calibration.

Wavelength calibration checks the difference between the wavelength measured by the instrument and the wavelength of the reference standard used. Specifically, different wavelength points can be calibrated across the instrument's entire range.

Microplate reader calibration

Technical scope for calibration of microplate readers

At Eyco we have technical expertise and standards recognized by ENAC, with guaranteed metrological traceability for the calibration of the wavelength scale and the detector scale in absorbance.

Measurement range
Range
Uncertainty (*)
--
Standard/Procedure
--
Instruments to be calibrated
Instruments
Dress
Code
TRANSMITTANCE OPTICAL TRANSMITTANCE DENSITY (ABSORBANCE)
OPTICAL TRANSMITTANCE DENSITY (ABSORBANCE)
400 nm < λ ≤ 780 nm
0,0 ≤ A < 1,0
1,0 ≤ A < 1,5
1,5 ≤ A < 2,0
2,0 ≤ A < 2,5
0,0065
0,025
0,055
0,065
PT39
Procedure
internal based on:
Royal Pharmacopoeia
Spanish
IUPAC Recommended
Reference Materials
for realization of
Physicochemical
properties
ELISA spectrophotometers
Microplate Readers
A,I
WAVELENGTH
Waveength
200nm ≤ ʎ ≤ 650nm0,30 nmPT39
Procedure
internal based on:
Royal Pharmacopoeia
Spanish
IUPAC Recommended
Reference Materials
for realization of
Physicochemical
properties
ELISA spectrophotometers
Microplate Readers
AI

The result

After the equipment calibration, you will receive a Calibration Certificate traceable to laboratories recognized by ENACThis certificate includes the full identification of your equipment, the calibration procedure followed, and the result.

Specifically, the result of the microplate reader calibration is a correction along with its uncertainty. The correction is the difference between the reference standard value and the value measured by your instrument. You will obtain the wavelength and absorbance corrections, as well as, if required, the stray light and linearity values.

Thus, based on the correction values ​​and their uncertainty, you can establish the acceptance and rejection values ​​of the equipment.

Also along with your equipment you will receive a calibration label with the certificate number, equipment identification and calibration date.

Frequently Asked Questions. Microplate Reader Calibration