UV-VIS spectrophotometry is used by a wide variety of entities and industries to control the quality of the products and services they offer.
The UV-VIS spectrophotometers are commonly used by a wide variety of industries for various purposes. From the food industry, companies and entities in charge of wastewater control, public health entities, companies and laboratories dedicated to the control of environmental qualityFrom the materials industry, many companies use UV-VIS spectrophotometry measurements to control the quality of their products or services.
Table of contents
Uses of UV-VIS spectrophotometry in industry
Food industry
If we focus on the food industryWe can find a wide variety of applications for this technique. It is common to use this equipment for quality control of food, by measuring the content of hydroxyproline, starch, or nitrites in meat and meat products; lactose in food and alcoholic beverages; or furosine in milk.
In addition, this technique is used to control trimethylamine, histamine, domoic acid or phosphorus in fishery and aquaculture products.
Likewise, it is common to use this equipment for measurements in fruit juices that allow control of citric acid, phosphorus, L-malic acid, D-isocitric acid, glucose, fructose and sucrose.
Similarly, in honey and bee products it is possible to measure hydroxymethylfurfural (HMF) with this technique.
It is also used UV-VIS spectrophotometry to determine volatile acidity, the amount of sulfur dioxide, acetic acid, L-malic acid, tartaric acid, D-gluconic acid or the intensity of color and hue in the wine.
In the olive industry, this technique is used to determine the so-called K Factor of olive oil, which measures the extinction coefficient of a fat and is directly related to the quality of the oil.
For its part, in animal feed, UV-visible spectrophotometry is used, for example, for the detection of urea or nitrites in feed.
Water control and treatment
Furthermore, the entities of public health and the companies responsible for the control and processing of waters are used UV-VIS spectrophotometers, among other things, for the detection of anionic detergents, cyanides, ammonium, nitrites, chromium or residual chlorine in drinking water and bottled water.
Environmental quality control
On the other hand, the use of this technique for the control of environmental quality It is commonly used by entities and laboratories that conduct tests to determine, for example, biochemical oxygen demand (COD), the amount of formaldehyde, anionic surfactants, phosphorus, phosphates, nitrites, nitrates, nitrogen, ammonium, chromium, phenols, or sulfides in inland waters. It is also commonly used to measure the amount of some of these compounds and chemical elements in wastewater and marine waters, or to determine the amount of Kjeldahl nitrogen in soils, sediments, and sludge.
Materials industry
But the use of the UV-VIS spectrophotometry It extends to many more industries, so it's possible to find applications of this technique in the materials industry as well. For example, a typical test for aluminum oxides and hydroxides is the determination of silica by UV-VIS spectrophotometry.
Textile industry
Other tests are carried out in the textile industrySome of these tests are, for example, the following: Cr (VI) by means of UV/VIS spectrophotometry in textile materials and finished textile articles, in metal accessories of textile garments, in paints and coating surfaces, in coated and laminated fabrics, or in plastic materials. Thus, the use of spectrophotometry is common for determining the ultraviolet protection factor (UPF) (for example, according to the Standards) UNE-EN 13758 or AS/NZ 43399); the calculation of solar and daylight characteristics in sun protection devices such as curtains (darkening, thermal and lighting comfort, etc.), among other things.
What is a spectrophotometer and how does it work?
The spectrophotometers are optical equipment that allow evaluating the light that is absorbed, transmitted or reflected by a material -solid or liquid- for each wavelength.
These equipments are made up of an illumination system, a monochromator and a detector.
Lighting system
The lighting system usually has a hydrogen or deuterium source for the ultraviolet spectrum area and a tungsten source for the visible and NIR spectrum area. In the case of UV-VIS spectrophotometers , they cover a range from 200nm to 900nm, which includes the ultraviolet, visible, and some infrared regions.
Monochrome
For its part, the monochromator makes it possible to separate the lighting spectrum and provide a beam of energy at a specific wavelength. Therefore, if the wavelength is critical or affects the measurements that are made, it is essential to calibrate this part of the equipment, or what is the same, perform a wavelength calibration, which will check the difference between the wavelength wave measured by the equipment and that of the reference pattern used.
Detector
Finally, the detector is responsible for transforming the signal it receives (photons) and converting it into an electrical signal. So, depending on the intensity of this signal, it can be recorded by a system that provides the numerical values of transmittance and absorbance or OD.
The calibration of the detector, also called the absorbance and transmittance scale, is essential to know the deviation of the equipment in the values of these magnitudes at each wavelength. The accuracy of these values is critical, since they allow comparing spectra of different materials or solutions with different concentrations. Specifically, the absorbance calibration is performed at all necessary wavelengths. To calibrate this scale, measurements of different absorbance values are usually made for each wavelength. Subsequently, the difference between the value measured by the equipment and the certified value of the standards used is evaluated.
What are calibration laboratories? Importance of an accredited calibration
The calibration laboratories accredited by ENAC offer a service that guarantees the results obtained. Our expert and highly trained staff, our facilities and equipment -suitable for carrying out calibrations- and the continuous assessment to which we are subjected by the accreditation entity ENAC In order to verify that we meet the necessary requirements for the reliability of our work results, our services are valued by a wide variety and number of companies and industrial sectors. Eyco Laboratory is accredited by ENAC to carry out calibrations in the wavelength scale and in absorbance or transmittance, both in portable equipment and in desktop equipment.
Scope of accreditation
Our scope of accreditation for these teams, in the scales mentioned above, is that which appears in the following table:
| RANGE Range | UNCERTAINTY (*) -- (*) | STANDARD/ PROCEDURE - Procedure | INSTRUMENTS TO CALIBRATE Instruments | CODE Code |
| TRANSMITTANCE | ||||
| (220-380)nm 0.01≤T<0.10 0.10≤T<0.30 0.30≤T≤1 |
0.050 0.060 T 0.020 T |
PT39 Internal Procedure based on: Royal Spanish Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | UV-VIS Spectrophotometers | A,I |
| (380-420)nm 0.00010≤T<0.0010 0.0010≤T<0.10 0.10≤T≤1 |
0.030 0.040 0.015 |
PT39 Internal procedure based on: Royal Spanish Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | ||
| (420-900)nm 0.00010≤T<0.0010 0.0010≤T<0.10 0.10≤T±1 |
0.30 0.080 0.0069 |
PT39 Internal procedure based on: Royal Spanish Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | ||
| TRANSMITTING OPTICAL DENSITY (ABSORBANCE) | ||||
| (220-380)nm 0.01≤A≤0.10 0.10≤A<0.50 0.50≤A≤2 |
0.0030 0.0060 0.0080 |
PT39 Internal procedure based on: Royal Spanish Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | UV-VIS Spectrophotometers | A,I |
| (380-420)nm 0≤A<1 1≤A≤3 3≤A≤4 |
0.060 0.050 0.15 |
PT39 Internal procedure based on: Royal Spanish Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | ||
| (420-900)nm 0≤A≤1.50 1.50≤A≤3 3≤A≤4 |
0.0040 0.050 0.20 |
PT39 Internal procedure based on: Royal Spanish Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | ||
| (900-2500)nm 0≤A≤0.8 0.8≤A≤2 |
0.0040 0.0090 |
PT39 Internal procedure based on: Royal Spanish Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | Spectrophotometers | A,I |
| WAVELENGTH | ||||
| 200nm≤λ≤850nm 850nm≤λ≤2000nm |
0.30 nm 0.50nm |
PT39 Internal procedure based on: Royal Spanish Pharmacopoeia IUPAC Recommended Reference Materials for realization of Physicochemical properties | UV-VIS Spectrophotometers | A,I |
Do you have doubts when interpreting your calibration certificate? We help you
The purpose of a measurement is to check whether a product or piece of equipment meets a given specification. That is, given some limits, which have been established as tolerated, it has to be determined if the values obtained for one or several characteristics of a product or equipment are within said limits.
This is why the uncertainty of any measurement process must be taken into account. This uncertainty depends on the methods of the measurement process, the equipment that has been used to carry out said measurement, the facilities and environmental conditions, the handling of the test items, human factors, etc.
A calibration certificate includes the results obtained in an exact, clear, objective and unambiguous way. The certificates must include the conditions under which the calibrations have been carried out, the uncertainty of the measurement and information regarding the evidence that the measurements are traceable.
In addition, information regarding the correction appears on the calibration certificates. This information is useful to correct the daily reading offered by a piece of equipment. Although the companies that request calibration certificates have qualified expert personnel to interpret certificates, it is possible that these personnel have doubts when interpreting them. That is why Laboratorios Eyco offers a personalized service to help our clients, who may require it, to make a correct interpretation of the certificate. Our expert staff makes available to our client companies the knowledge and experience available to translate the results that may raise doubts.
FAQ
HTMLWhat is UV-VIS spectrophotometry and what is it used for?
La UV-VIS spectrophotometry It is a quantitative analytical technique that measures the amount of ultraviolet and visible radiation absorbed by a substance as a function of wavelength. It is used to identify compounds, determine their concentration in a solution, and analyze the optical properties of various materials and products in industry.
How does a UV-VIS spectrophotometer work?
Un UV-VIS spectrophotometer It works by passing a beam of light through a sample. The equipment consists of a light source, a monochromator (which selects the specific wavelength), a sample compartment, and a detector. The detector measures the intensity of the light before and after it passes through the sample, thus calculating the absorbance or transmittance of the substance.
What are the key applications of UV-VIS spectrophotometry in industry?
This technique is fundamental in numerous sectors. It stands out in the Quality Control of chemical and pharmaceutical products, water analysis, characterization of optical and plastic materials, monitoring of biotechnological processes and determination of compounds in the food and beverage industry (such as the measurement of color and polyphenols).
What are the differences between single-beam and double-beam spectrophotometers?
The main difference lies in their optical design and precision:
- Teams keep it simple They measure the reference (blank) and the sample sequentially, requiring manual adjustment or prior calibration that can be affected by drift in the light source.
- Those of double beam They split the light simultaneously towards the sample and the reference, automatically compensating for lamp fluctuations and offering greater stability and reproducibility in continuous measurements.
Why is it necessary to calibrate a spectrophotometer periodically?
La periodic calibration In a specialized laboratory like Eyco, it is essential to ensure that absorbance, transmittance, and wavelength readings remain accurate and traceable to international standards. Over time, wear and tear on optical lamps and mechanical components can alter the results, compromising the technical rigor of industrial tests.