The standards of the main pharmacopoeias in relation to spectrophotometry focus on the control of wavelength, absorbance, photometric linearity, stray light limit and resolution.
The main pharmacopoeias of the world establish a series of standards or requirements related to the calibration of spectrophotometersThus, for example, parameters such as wavelength accuracy, absorbance, and photometric linearity must be regularly monitored. Similarly, pharmacopoeias establish requirements for monitoring stray light limits and resolution.
Table of contents
What is a pharmacopoeia?
A pharmacopoeia is a legislative framework that includes both the most common medicinal substances and the quality standards or requirements for preparing and combining them.
Among the most recognized pharmacopoeias in the world are the United States Pharmacopeia (USP) and the European Pharmacopoeia (EP).
Main pharmacopoeias and their standards in relation to spectrophotometry
North American pharmacopoeias y European They establish a series of requirements for the control of measurements that are carried out by spectrophotometers.
Thus, for example, pharmaceutical laboratories must ensure that measurements related to wavelength or absorbance are accurate. Similarly, the control of photometric linearity, stray light limit, and resolution must be performed regularly.
Below are the requirements that establish the USP and EP thereon.
North American Pharmacopeia (USP) Standards
Following the revision of USP chapter 857, it was updated in December 2019 and the 42nd edition of the United States Pharmacopeia.
Chapter 857 (USP 42 NF 37) establishes, in the first place, as a new requirement that the control of the UV-VIS spectrophotometer depends on the operating range (absorption range and wavelength). Secondly, from the update of this chapter, the determination of the photometric linearity is mandatory. Finally, the new edition introduces some changes and clarifies discussion points.
Below are the changes introduced in the new update (857).
As far as wavelength accuracy is concerned, prior to the last update, the USP stated:
-First of all, to check the precision of the wavelength from 240nm to 650nm, a holmium filter must be used.
-Second, to check wavelength accuracy above 650nm (730nm to 870nm), a didymium filter must be used.
Furthermore, the latest edition of the USP (857) states that:
-For wavelength accuracy below 240nm, spectrophotometers should be checked with a cerium filter (200nm to 255nm).
As far as absorbance precision control is concerned, prior to the last update, the USP had established that:
-In the UV range, potassium dichromate solutions are used in the intended operating range up to 3 Abs.
-In the visible range, neutral density filters are used in the intended operating range up to 3 Abs.
In addition, after the update (857) it is indicated that:
-In the UV 0-1 Abs range, potassium dichromate solutions with concentrations of 20-60 mg/l, or nicotinic acid solutions with concentrations of 6-24 mg/l, can be used.
-In the range of UV 1-3 Abs, potassium dichromate solutions with concentrations of 80 to 200 mg/l can be used.
Regarding the control of photometric linearity, the update of the USP (857) explains that:
"Verification of photometric response (linearity) is required and should be assessed using a type of standard appropriate for the required wavelengths, where at least three different absorbance levels appropriate and spanning the required operating range are measured."
In addition, the USP suggests the following filters or reference materials:
-In the UV range, 6-60mg/l nicotinic acid solutions; and potassium dichromate solutions 20-200 mg/l
-In the visible range, neutral density glass filters.
On the other hand, the USP indicates that at least three are necessary to verify photometric linearity. filters with different concentrations per spectral range. Furthermore, it is not necessary to calculate the correlation coefficient of the standard response to demonstrate linearity. Thus, it is sufficient to demonstrate that the acceptance limits have been met at the different absorbance levels selected.
As regards the limit of stray light, it should be said that prior to the last update, the USP It stipulated that stray light levels must be controlled in the 200-400nm wavelength range.
To measure stray light, a 5mm reference filter and a 10mm filter are required, using the same solution.
Furthermore, following the update of USP (857), two possible procedures are described for verifying the stray light limit:
-The first procedure (procedure A), coincides with the previously described method.
The second procedure (procedure B) states that the stray light filter can be measured with a path length of 10 mm against a reference filter filled with pure water. However, acetone is an exception, and its absorbance is measured against air. The absorbance value measured by the filter must be greater than 2.0 at the recommended wavelength indicated in the table.
| Filter/Solution | Spectral range | Recommended wavelength |
| Potassium chloride (12g/l) | 190-210nm | 198 nm |
| Sodium iodide (10g/l) | 210-270nm | 220 nm |
| Potassium iodide (10g/l) | 210-270nm | 220 nm |
| Acetone (pure) | 250-330nm | 300 nm |
| Sodium nitrite (50g/l) | 300-400nm | 340nm |
Regarding the resolution of the spectrophotometer, prior to the last edition, the USP It stipulated that, to verify it, a liquid toluene filter in hexane and a hexane reference filter must be used. In this respect, there have been no modifications following the update of USP (857).
European Pharmacopoeia (EP) Standards
As for the latest version of the European Pharmacopoeia (EP 10), which has been in effect since January 2020, includes a comprehensive revision of section 2.2.25 (Ultraviolet and Visible Absorption Spectrophotometry). Thus, according to the latest revision, the control of UV-VIS spectrophotometers compliant with EP 10 depends on the intended spectral range (absorption range and wavelength). This means that the filters Standard filters can only be partially used to verify parameters. Therefore, these filters must be individually supplemented according to the intended spectral range.
The following sections show the new requirements of EP 10 with regard to chapter 2.2.25.
Regarding this parameter, before the last revision, the EP 8 stated that the holmium filter is to be used to verify the accuracy of the wavelength from 240nm to 650nm.
Furthermore, following the latest amendment, EP 10 requires verifying the wavelength accuracy of an appropriate number of bands within the intended spectral range using one or more reference materials. Specifically, it is recommended to test at least two wavelengths covering the intended spectral range.
Regarding reference materials (filters), EP 10 suggests the following:
-Use a liquid holmium filter for a spectral range of 240nm to 640nm.
-Make use of a didymium liquid filter for a spectral range greater than 640nm.
-Use a liquid cerium filter for a spectral range below 240nm.
| Material | Wavelength (nm) |
| Cerium liquid filter | 201, 211, 223, 240, 254 |
| Holmium Liquid Filter | 512, 732, 794 |
| Holmium Liquid Filter | 241, 287, 361, 451, 485, 537, 640 |
Regarding the precision of the absorbance, it was established in the previous version of the EP (EP8) that a 60 mg/l potassium dichromate solution should be used for the UV range; while for the VIS range, the potassium dichromate solution to be used was 600 mg/l.
However, following the latest revision, EP 10 further states that control of the accuracy of absorbance is required at an appropriate number of wavelengths in the intended spectral range using filters solids or liquids. In addition, EP 10 specifies that at least approximate values must be verified at the two limits of the absorbance range.
Regarding the suggested reference materials (filters), it is specified that:
-Use a solution of nicotinic acid (Niacin) up to 2.0 Abs (5-45 mg/l), for the UV range.
-Neutral density glass filters up to 2.0 Abs, are used for the VIS range.
| Material | Wavelength (nm) |
| Niacin liquid filter (5-45 mg/l) | 213.61 |
| Neutral Density Glass Filter | 440, 465, 546, 1, 590, 635 |
Even so, it is still possible to use potassium dichromate filters to verify absorbance accuracy in the UV range.
As regards the photometric linearity, the EP Article 10 states that it must be controlled within the specified spectral range. Furthermore, it suggests the following reference materials (filters):
-Nicotinic acid solution (Niacin) for the UV range.
-Neutral density glass filters for the VIS range.
| Material | Wavelength (nm) |
| Niacin liquid filters (5-40 mg/l) | 213, 261 |
| Neutral Density Glass Filters | 440, 465, 546.1, 590, 635 |
On the other hand, it is important to point out that to check the photometric linearity at least three are required filters with different concentrations by spectral range.
As for the limit of stray light, the EP 10 requires its control at a suitable wavelength. The acceptance criterion depends on the filter used.
Specifically, EP 10 suggests the reference materials (filters) indicated in the table.
| Filter/Solution | Absorbance at wavelength |
| Potassium chloride (12g/l) | ≥ 2.0 Abs at 198nm |
| Sodium iodide (10 g/l) | ≥ 3.0 Abs at 220nm |
| Potassium iodide (10 g/l) | ≥ 3.0 Abs at 250nm |
| Sodium nitrite (50 g/l) | ≥ 3.0 Abs at 340nm and 370nm |
As far as resolution control is concerned, the EP 10 states perform it with toluene in hexane and use hexane as the reference solution.
Calibration of equipment and filters to meet the standards of the main pharmacopoeias
Periodically check that the equipment y filters are properly calibrated to meet the standards of the main pharmacopoeias, it is essential. In this way, good measurement results can be ensured.
Therefore, it is necessary to have accredited calibration laboratoriess by ENACSince laboratories like Eyco provide a guarantee of the service they offer due to the continuous evaluation to which we are subjected. In other words, through these evaluations by ENAC, our laboratory can demonstrate that it has the necessary accreditation. technically trained expert staff for carrying out the calibration of equipment and filters, as well as with the adequate facilities and equipment to perform these calibrations.
Furthermore, thanks to this continuous evaluation by ENAC, accredited laboratories must adapt their processes to meet the required standards and, therefore, offer reliable results to our clients.
Laboratorios Eyco, experts in optical calibrations
Eyco Laboratories has the most extensive scope of accreditation in Spain in the optical area. Thus, our laboratory is accredited to calibrate both UV-VIS spectrophotometers , the filters, according to the standards of the main world pharmacopoeias.
Firstly, regarding the calibration of UV-VIS spectrophotometers, our laboratory is accredited by ENAC to perform optical calibrations on the wavelength scale and in absorbance or transmittance, both in portable and desktop equipment.
Furthermore, regarding filters, Eyco Laboratories also performs calibrations according to the standards of the main pharmacopoeias. For example, we are accredited to calibrate wavelength filters, absorbance and transmittance filters, linearity filters, and scatter filters.
For more extensive information, you can consult our scope of accreditation here.
FAQ
Why is it important to measure the color of wine?
El wine color It is one of the main indicators of quality, age, and state of preservation. Its measurement allows for the evaluation of the wine's evolution during the production and aging process, the detection of possible alterations, and the assurance that the product maintains the characteristics expected by the consumer.
What is UV-Vis spectrophotometry applied to wine analysis?
La UV-Vis spectrophotometry It is an analytical technique that measures how a liquid absorbs light at different wavelengths. In the case of wine, it allows for the objective quantification of its hues and color intensity by measuring absorbance at specific wavelengths of the visible spectrum.
What wavelengths are used to analyze the color of wine?
They are commonly used specific wavelengths of the visible spectrum to determine the intensity and hue of the wine. These measurements allow the calculation of parameters such as color intensity or shade, which are fundamental for evaluating red, rosé, and white wines objectively and reproducibly.
What official methods exist for measuring the color of wine?
Several standardized methods Used in oenological laboratories to measure the color of wine. Among the best known are:
- The color system CIE, which describes color using chromaticity coordinates.
- The method of Sudraud, used to determine color intensity.
- The indices of Glories, which analyze the contribution of different phenolic compounds.
- The methods defined by the OIV (International Organisation of Vine and Wine).
Why have wine color measured in an accredited laboratory?
Perform the analysis in a laboratory accredited by ENACAs Eyco, accreditation guarantees reliable, traceable, and repeatable results. It ensures that the analytical methods used meet international standards and that the equipment is properly calibrated, providing accurate data for wineries, research centers, and oenological quality control.